SPECIFICATIONS
| Model | IT-550F | IT-550L | IT-550S |
| Measuring range | -50 to 500°C | ||
| Spectral response | 8 to 16 mm |
||
| Sensor & lense | Thermopile, Silicon | ||
| Display range (resolution) |
-55.0~505.0°C
(0.1°C) |
-55
~ 505°C (1°C) or -55 ~ 505.0°C (0.1°C) |
|
| Accuracy
(Conditions:
case temperature 18~28°C, relative humidity 55%, e=1.00) |
-50.0
~ -0.1°C: Within ±(10% of rdg -2)°C 0.0~200.0°C:
Within ±2.0°C 200.0~500.0°C:
Within ±(1% of rdg)°C |
-50 ~ -1°C / -50.0 ~ -0.1°C: Withhin ±(10% of rdg -2)°C 0~200°C / 0.0~200.0°C: Within ±2°C / ±2.0°C 200~500°C
/ 200.0~500.0°C: Within ±(1% of rdg)°C |
|
| Repeatability | -50.0~-0.1°C: Within ±1.0°C 0.0~500.0°C:
Within ±0.5°C |
Resolution
0.1°C: -50.0~-0.1°C:Within ±1.0°C 0.0~500.0°C: Within ±0.5°C
Resolution 1°C:-50~-1°C: ±2°C 0~-500°C: ±1°C |
|
| Response (95% response) | Less than 1.6s |
Less
than 0.7s (resolution 1°C) Less than 1.6s (resolution 0.1°C) |
|
| Field of View |
ø
24±3 mm at 1000 mm |
ø
2.5 mm at 70mm |
|
| Emissivity |
Adjustable
in 0.10~1.00, in 0.01 increment |
||
| Sighting |
Twin
laser beam marker (class II) |
||
| Protection |
Splash-proof,
IP54(IEC529) |
||
| Operating range |
0~40°C,
relative humidity 35~85%, and no dew condensation |
||
| Power source |
9V
battery (6F22 or 6LR61) x 1 or AC adapter |
||
| Battery life |
20
continuous hours with laser marker in use (alkaline battery) |
||
| Dimension & weight |
L200
x W47 x H48 mm, 280g (battery included) |
||
| Functions | Data
memory (130 points), printer output, RS232C, auto power off, camera tripod
screw hole |
Alternative
display resolution, analog output(0~1VDC scalable), RS232C, Max/Min value
display, auto power off, camera tripod screw hole |
|
| Storage temperature |
-20~55°C,
no dew condensation |
||
| Standard accessories | instruction manual, carrying case, hand strap, battery, screw-driver | ||
Note: Use of analog output, printer output, RS-232C, and the AC adapter requires the expansion kit option (sold separately
APPLICATIONS
|
|
||
|
IT-550F ( For Field Use) Temperature management for food items and preventive maintenance for power supplies. |
IT-550L (For Laboratory Use) Temperature measurement for laboratory experiments. |
IT-550S (For Laboratory Use Sport Focus Model) Temperature measurement for IC surfaces and other small areas. |
FIELD OF VIEW
|
IT-550S: ø 2.5mm at 70mm
|
|
IT550F & IT550L: ø 24mm at 1000mm
|